DN80 Flow Range
The DN80 inline vortex meter operates within a broad flow envelope. For liquid service, the usable range runs from 12 m3/h at the low end to 130 m3/h at the top, equivalent to roughly 53 to 572 US gallons per hour. Gas service expands that considerably: 80 to 800 m3/h (approximately 2,825 to 28,250 SCFH). Both figures assume standard Reynolds number conditions between 2×10^4 and 7×10^6.
Turndown is 1:6 to 1:30 depending on fluid type and process conditions. That wider end of the turndown range is achievable on clean, low-viscosity liquids. Steam and gas applications typically land in the 1:10 to 1:15 zone in practice.

One thing worth clarifying: for liquid measurement, no temperature or pressure compensation is needed. The vortex shedding frequency is directly proportional to volumetric flow rate, and the meter outputs that as a pulse or 4-20 mA signal without correction. For gas and steam, the built-in temperature and pressure compensation option (BPT code) gives you mass flow or standard volumetric flow without needing an external flow computer.
| Parameter | DN80 Specification |
| Liquid Flow Range | 12 ~ 130 m3/h (53 ~ 572 US gal/h) |
| Gas Flow Range | 80 ~ 800 m3/h (2,825 ~ 28,250 SCFH) |
| Liquid Velocity | 0.5 ~ 7 m/s |
| Gas / Steam Velocity | 5 ~ 70 m/s |
| Accuracy (Liquid) | +/- 1.0% |
| Accuracy (Gas/Steam) | +/- 1.5% |
| Repeatability (Liquid) | +/- 0.33% |
| Repeatability (Gas/Steam) | +/- 0.5% |
| Turndown Ratio | 1:6 ~ 1:30 |
| Reynolds Number | 2×10^4 ~ 7×10^6 |
DN80 Dimensions
The STLU DN80 is available in wafer and flanged configurations. Dimensions below are extracted from the Silver Instruments catalog. Note that the C2 dimension for the wafer type is listed as 800 mm, which is longer than the standard compact units at DN80; this applies to the remote transmitter version with an extended neck. The C1 dimension is the standard compact housing height.
Wafer Connection (DN80)

| DN | A (mm) | B (mm) | C1 (mm) | C2 /(mm) | CH |
| 80 | 84 | 118 | 440 | 500 | 800 |
Flange Connection (DN80)

| DN | A (mm) | B (mm) | C1 (mm) | C2 (mm) | Bolt | Bolt Qty |
| 80 | 180 | 160 | 490 | 550 | M16 | 8 |
Images showing both wafer and flange configurations are available on the Silver Instruments product page and will be inserted here.
What Does a DN80 Vortex Meter Typically Measure?
The 80 mm bore sits in a practical middle ground: large enough to handle meaningful industrial flow volumes, small enough to keep installation compact. Because of that, DN80 vortex meters appear across a surprisingly wide range of applications.
DN80 Steam Flowmeter— The Most Common Application

DN80 Vortex flow meter for steam
Steam is the single most common fluid we see at DN80. Saturated steam in the 0.3 to 1.0 MPa range, delivered through 3-inch headers in food plants, breweries, pharmaceutical clean rooms, hospital HVAC systems, and small power stations. The vortex meter handles it well because there are no moving parts to erode, the piezocrystal sensing element sits sealed inside the bluff body and does not contact the steam directly, and the high-temperature option (extended range up to 300 degrees C) covers superheated steam conditions.
For steam, we almost always configure the meter with built-in temperature and pressure compensation (BPT). That eliminates the separate flow computer and gives the customer a direct mass flow or steam totalizer output on a single instrument. Last year a food manufacturer in Southeast Asia switched from an orifice plate to our DN80 STLU on a 0.8 MPa steam header. The pressure drop reduced significantly, and the plant now reads steam mass flow directly on the local display without any external calculation.
DN80 Compressed Air and Industrial Gas Flow meter

DN80 Vortex flow meter for compressed air
Compressed air distribution is the second-most frequent application. Factory air systems at 6 to 10 bar, branch lines running 80 mm pipe, where the goal is monitoring consumption by department or machine. Vortex works here because compressed air is clean, dry, and above the minimum Reynolds number at typical operating velocities.
Nitrogen is also common, particularly in chemical plants and electronics manufacturing where purge and blanketing flows need monitoring. Natural gas in 3-inch meter runs for small industrial burners and process heaters falls in this bore range too. For all gas applications, the BPT compensation option applies, and the RS-485 Modbus output connects directly to a plant DCS or SCADA without a gateway.
DN 80 Water and Cooling Fluids Flow meter

Chilled water and cooling water loops in HVAC and process cooling systems frequently use DN80 vortex meters. The fluid is clean, the temperatures are moderate (typically 5 to 45 degrees C), and there is no chemical compatibility concern with 304 SST wetted parts. The 12 to 130 m3/h liquid range covers most branch cooling circuits in medium-sized industrial plants.
Hot water circuits in district heating, boiler feed preheating loops, and heat exchanger monitoring are also good fits. Vortex meters do not require straight pipe runs as long as some other technologies, but we still recommend 10 to 15 diameters upstream and 5 diameters downstream for best accuracy.
DN80 Thermal Oil and Heat Transfer Fluids Flow meter
Thermal oil systems operating at 150 to 280 degrees C (Therminol, Dowtherm, and similar products) show up frequently in plastics processing, rubber vulcanization, and chemical batch reactors. The DN80 extended temperature option handles this range, and 316 SST wetted parts (M1 material code) are specified when the heat transfer fluid has any chloride content or acidic additive packages.
Viscosity is worth checking here. Thermal oil viscosity at startup temperatures can exceed 50 cP, which pushes the Reynolds number below the minimum threshold and gives unreliable readings until the system warms up. This is not unique to the vortex meter: any velocity-based meter faces the same issue. Most thermal oil customers accept a warm-up period of 10 to 20 minutes before relying on the flow reading.
DN80 Light Chemical Fluids Flow meter
Solvents, alcohols, acetone, and dilute acid or caustic solutions in chemical plant branch lines run through DN80 vortex meters where the process temperature and viscosity are within spec. For anything above a mild acid or caustic concentration, we check the liner and electrode compatibility first. The STLU does not use a liner in the same way an electromagnetic meter does, but the 316 SST bluff body and process connection material need to match the fluid.
Industries Where DN80 Vortex Meters Are Commonly Installed
- Power and utilities: steam distribution to turbines, auxiliary systems, and deaerators.
- Food and beverage: steam to autoclaves, CIP hot water circuits, and cooling water to fermentation tanks.
- Pharmaceutical: clean steam and WFI distribution, process gas flows in formulation areas.
- Chemical processing: nitrogen blanketing, solvent recovery, heat transfer oil.
- HVAC and building utilities: chilled water, hot water, compressed air sub-metering in commercial and industrial buildings.
- Plastics and rubber: thermal oil in extruder and mold temperature control systems.
- Marine and offshore: compressed air utility systems, seawater cooling secondary circuits (316 SST specified).
Why Choose Vortex Over Other Technologies at DN80?
At 80 mm, the main competing technologies are electromagnetic meters, differential pressure (orifice plate), and ultrasonic clamp-on.
Electromagnetic meters are excellent for conductive liquids but cannot measure steam, compressed air, or gas at any flow rate. For a site that monitors both steam and cooling water, a vortex meter can handle both fluids on a single platform. That simplifies spare parts and maintenance training.
Orifice plates are still installed in many plants, but the pressure drop penalty at 80 mm is real. Vortex meters generate roughly one-quarter to one-half the permanent pressure loss of a comparable orifice plate. Over a year of operation on a continuous steam or compressed air system, that pressure drop difference translates to measurable energy cost.
Clamp-on ultrasonic works well for liquids and requires no process penetration, but it does not perform on steam or gas at all, and accuracy degrades on smaller bores. DN80 is right at the lower edge of reliable clamp-on operation.
Model Selection for DN80 Applications
The Silver Instruments STLU model code for a standard DN80 flanged steam meter with built-in T+P compensation, HART output, and explosion-proof housing would look like this:
STLU – VFN – 1 – 4 – 08 – D – 1 – 4 – 1 – G1 – d – 0 – BPT – M0
Breaking that down: VFN = vortex meter, 1 = flanged connection, 4 = steam, 08 = DN80, D = with display, 1 = 24V DC power, 4 = HART output, 1 = standard temperature range to 250 degrees C, G1 = GB 1.6 MPa flange standard, d = flameproof explosion proof, 0 = IP65, BPT = built-in temperature and pressure compensation, M0 = 304 SST wetted parts.
For cooling water or chilled water, swap fluid code to 2 (liquid) and compensation to B0 (none needed). For nitrogen or compressed air, fluid code 3, and BPT compensation.
Application Note: Steam Sub-Metering at a Beverage Plant in Malaysia
A beverage manufacturer in Malaysia needed to track steam consumption by production line. They had four DN80 branch headers coming off a main 6-inch steam header, all running saturated steam at around 0.6 MPa and 165 degrees C.
We supplied four STLU DN80 flanged vortex meters with the BPT built-in compensation, HART 7 output, and flameproof housing. The instruments connect to a plant-level energy management system via a Modbus gateway. Steam mass flow and daily totalized consumption are now visible per line, which allows the production team to identify leaks or overconsumption at a branch level instead of hunting through the whole system.
The entire installation took about one day per meter. The main adjustment was verifying that minimum upstream straight pipe of 10D was available on each branch. Two of the four branches needed a short spool piece to meet that requirement.
Frequently Asked Questions
Vortex Flow Meter Products
Request a Quote for DN80 Vortex Flow Meter
To get a specific product recommendation and price, send us the following details:
- Pipe size confirmation: DN80 (3 inch)
- Fluid type: steam / compressed air / nitrogen / water / thermal oil / other
- Operating pressure (bar or MPa)
- Operating temperature (degrees C)
- Flow range: minimum and maximum (m3/h or kg/h for steam)
- Required output signal: 4-20 mA, HART, Modbus RS-485, or pulse
- Process connection standard: ANSI, DIN, or GB flange
- Explosion-proof requirement: Zone 1, Zone 2, or non-Ex
- Hazardous area classification if applicable
Email your specification to sales@silverinstruments.com or visit silverinstruments.com to see the full STLU vortex flow meter range.